2525 Series Drop-Through Static Load Cell (10 N)
2525-808 | 2525 Series Drop-Through Static Load Cell (10 N)
Static Load Cell: 10 N (1 kg, 2 lb), Type OOf, Of
Tension/Compression
Type OO female fitting (6 mm connection with 2.5 mm diameter clevis pin)
Includes adapter to Type O female fitting (12 mm connection with 6 mm diameter clevis pin)
- Drop through, bolt on design with self identification and electrical calibration
- Linearity: +/- 0.25% of reading over a range of 0.4% to 100% of capacity
- Repeatability: +/- 0.25% of reading over a range of 0.4% to 100% of capacity
- Hysteresis: 0.2% of maximum reading in cycle
- Creep: 0.2% of reading over 3 minutes
- Output Sensitivity: 2.0 millivolts per volt +/- 20% at full rated capacity
- Bridge Resistance: 350 ohms +/- 5%
- Overload Protection: 150% of capacity without permanent zero shift. 300% of capacity or 5 kN (500 kg, 1000 lb) whichever is greater without mechanical failure
- Zero Stability: 0.001% of full rated output per hour (isothermal conditions)
- Temperature Compensation: 0 to 50 degrees Celsius (32 to 122 degrees Fahrenheit)
- Temperature Effect on Zero and Span: 0.002% of capacity per degree Celsius.
Diamètre | 80 mm (3.1 in) |
---|---|
Capacité | 0.01 kN (2 lbf) |
Montage mécanique | 2.5 mm clevis pin (Type OOf) |
Static or Fatigue Rated | Static |
Upper Effective Length | 135 mm (5.3 in) |
2525-800 Series Load Cells
Rated capacities from ±2.5 N to ± 100 kN (0.25 kgf to 10,000 kgf, 0.5 lbf to 22,500 lbf) ISO DIS 7500/1, EN10002-2, BS1610 (1992) and JIS B7721, B7733
- Produits
- 4/29/2005
- 128.3 KB